HPLC Calibration Procedure | High performance Liquid Chromatography Calibration

Rahul Kashyap
0

Purpose:

The purpose of this SOP is to provide written guidelines for the Calibration of High
performance Liquid Chromatography
Calibration of Pump :
Flow Test :
Check the flow at the out let of the detector with column in a standard   
measuring   cylinder to verify the flow.
Connect C18 column with 0.45  filtered Distilled water as mobile phase.
Check the flow for 0.5 ml/min, 1.5ml/min & 2ml/min.
Limit : ± 1%
Frequency: Once in a month.

Pressure Linearity :
Check the pressure at 0.5 ml/min in Kgf  and then check the corresponding pressure at 1.0, 1.5, 2.0 and 2.5 ml/min and observe the pressure in Kgf.
Limit : ± 10 % of actual pressure. 
Frequency: Once in a month.

System Suitability :

(Ref : IP 1996)
Check for relative standard deviation on daily basis/as & when required on
three replicate injections. (as per BP) (Limit NMT 2%)
Determination of Column Efficiency
Purge the HPLC system with solvent like distilled water  to remove the traces of impurities present in the tubing & Connecting parts, if any.  Cleaning of injection port & syringe is also necessary. Make sure that the whole system is perfectly cleaned.
Connect the column with the direction of flow arrow pointing towards detector, if mentioned on column. For a new column, first give slow washings by solvents like Acetonitrile, Methanol (HPLC Grade) & distilled water filtered through 0.2  membrane filter. Flow rate is increased gradually from 0.5ml/min. up to 1.5 ml/min. increments. Thus equilibrate the column when it is 1st installed.
 
Determination of Column Efficiency 
Where
                       
                        X             =          Arithmetic mean of set
                        Xi            =          Individual measurement in a set of N measurement

Determination of Column Efficiency :

Perform an efficiency test on the new column by following procedure :
Prepare mobile phase with the combination of 70:30 (Methanol & water). Prepare mobile phase freshly on the day of use. Degas the mobile phase.
Chromatographic conditions :
              = 254 nm 
Test Solution - In a 50 ml volumetric flask, take 1.0 ml of Benzene then volume make up to mark with the methanol. Pipette out 5.0 ml in 50 ml volumetric flask and volume make upto the mark with methanol.
Equilibrate the column with mobile phase till the detector displays constant absorbance or Zero absorbance. Column saturation can be checked 00. 0 Pump absorbance should be zero.
1st check the absence of impurities in mobile phase by injection it through the column. A straight line printed on chromatogram shows
that there is no any impurity present in mobile phase or in the column. i.e. stationary phase.
Now inject the filtered test solution after satisfactory separation of the components from mix. Chromatogram draws the straight line passing through the peak center. Measure the width of the peak at it's half height. Generally this calculations are done on middle peak, i.e. benzene peak
Calculate the no. of theoretical plates (N) by using formula:
                   Where,
                                    N         =          16 x [T/W}2
                                   16        =          constant
                                    T         =          Retention time of the peak of interest.
                                    W        =          Width of peak at it's 5.0 % height.
                                     
Generally theoretical are plates calculated as plates/meter.
Wash the column with water, acetonitrile & methanol so as to clean the column free from any traces of test solution. Flushing with a heat organic solvent is usually sufficient to remove any contaminant. Store the column in a suitable solvent, which is prescribed by manufacture, or in a solvent like 60:40 (methanol & water)
Theoretical plates can be calculated periodically or when required. After the use of column, make sure that the end fittings are firmly in place which prevents column drying.
Frequency : Initially when a new column is purchased or Yearly required.
Calibration of the Detector:
Check the reference energy of the dectetor  lamp. The energy of the D2 lamp should not be less than 730 at 230 nm.
Linearity of the Detector Response, and Auto injector. and
Uniformity of the Auto Injector   is done as follows.
Chromatographic condition as follows
Column             : ODS C18 25cm X 4.6 mm
Mobile Phase   : Methanol : Water(70:30)v/v
Flow Rate         : 1.0 ml /min.
Wavelength      : 254.0 nm.
Injection Qty.    : 20 ul.
Prepare 3 different ratios of benzene and tolune in methanol as given below and inject each solution in triplicate.
Standard Preparation.:
LEVEL 01
Pipette out 0.5 ml of benzene and 1.0 ml of tolune into 50 ml V.F. and dilute with methanol upto mark . Further dilute 5.0 ml to 50 ml with methanol.
LEVEL 02
Pipette out 1.0 ml of benzene and 1.0 ml of tolune into 50 ml V.F. and dilute with methanol upto mark . Further dilute 5.0 ml to 50 ml with methanol.
LEVEL 03
Pipette out 1.5 ml of benzene and 1.0 ml of tolune into 50 ml V.F. and dilute with methanol upto mark . Further dilute 5.0 ml to 50 ml with methanol.
The retention time of benzene & tolune should be reproducible in each three levels
{Limit ±0.5 mins.}
Calculate the area ratio of benzene to tolune . And Calculate RSD of each level is NMT 1.0 % . And plot the graph of  a) Level verses Area Ratio. B) Peak Area of
Benzene Verses Concentration % v/v C)Peak Area of Benzene verses  no. of injection. D) Peak Area of Tolune verses No. of injection.
Frequency : Yearly required.
Precautions:
Wash the column thoroughly before and after use with prescribed solution .
Wash the injection port with water and with prescribed mobile phase before 
use.
Inject the sample only after the system is stabilized to get accurate results.
Close the instrument as per the sequence described in above procedure.
Maintain the cleanliness of instrument before & after use at each time.

Post a Comment

0Comments
Post a Comment (0)